Accumath Toolchest — Thermocouple Reference Tables (NIST ITS-90)

Data file: accumath-thermocouple-tables.csv Columns: type, temp_C, emf_mV, source Rows: 1,211 data rows (plus header)

Basis and conventions

Coverage

Type Class NIST range (°C) Rows EMF at range ends (mV)
K base metal −270 to 1372 166 −6.458 to 54.886
J base metal −210 to 1200 142 −8.095 to 69.553
T base metal −270 to 400 68 −6.258 to 20.872
E base metal −270 to 1000 128 −9.835 to 76.373
N base metal −270 to 1300 158 −4.345 to 47.513
R noble metal −50 to 1768.1 183 −0.226 to 21.101 (at 1768 °C)
S noble metal −50 to 1768.1 183 −0.236 to 18.693 (at 1768 °C)
B noble metal 0 to 1820 183 0.000 to 13.820

Endpoint convention: the grid is every 10 °C across each type's full NIST range. Type K's upper limit, 1372 °C, is off the 10 °C grid and is included as a final row. Types R and S are defined to 1768.1 °C; their last row is 1768 °C, the last integer-degree point of the NIST table (21.101 mV for R, 18.693 mV for S).

How the values were produced (route used)

The CSV values were computed from NIST's own published forward reference-function polynomials — route 2 of the build plan — and then verified against NIST's published tables at every single grid point.

For each type, NIST publishes a polynomial reference function E(t) = Σ cᵢ tⁱ (type K above 0 °C adds an exponential term a₀·exp(a₁(t − a₂)²)). These polynomials are not approximations of the tables; they are the defining reference functions of Monograph 175, and NIST's printed tables are generated from them. The forward coefficients were transcribed from the coefficient section of each type's file in the NIST SRD 60 distribution:

No third-party coefficient set (Omega or otherwise) was used; no value came from memory.

Polynomial subranges (°C)

Type Subranges
K −270–0; 0–1372 (with exponential term)
J −210–760; 760–1200
T −270–0; 0–400
E −270–0; 0–1000
N −270–0; 0–1300
R −50–1064.18; 1064.18–1664.5; 1664.5–1768.1
S −50–1064.18; 1064.18–1664.5; 1664.5–1768.1
B 0–630.615; 630.615–1820

Verification against the published NIST tables

The same SRD 60 files contain NIST's published 1 °C tables. The value at every 10 °C grid point of those tables (1,211 points) was compared with the computed value:

Representative spot checks (computed = published):

Type t (°C) EMF (mV) Type t (°C) EMF (mV)
K −270 −6.458 N −270 −4.345
K −100 −3.554 N −100 −2.407
K 100 4.096 N 100 2.774
K 500 20.644 N 600 20.613
K 1000 41.276 N 1300 47.513
K 1372 54.886 R −50 −0.226
J −210 −8.095 R 100 0.647
J −100 −4.633 R 1000 10.506
J 100 5.269 R 1768 21.101
J 760 42.919 S −50 −0.236
J 1200 69.553 S 100 0.646
T −270 −6.258 S 1000 9.587
T −200 −5.603 S 1768 18.693
T 100 4.279 B 0 0.000
T 400 20.872 B 100 0.033
E −270 −9.835 B 500 1.242
E −100 −5.237 B 1000 4.834
E 100 6.319 B 1820 13.820
E 1000 76.373

Accuracy note — what these tables are and are not